| Functions: |
 |
Filling |
 |
Checkweighing |
 |
Percentage |
 |
Statistics |
 |
Formulation |
 |
GLP
procedures |
 |
Infrared
sensors |
|
 |
3
years warranty period |
| |
|
|
| Technical data: |
MYA 0,8/3/2Y |
MYA 5/2Y/F |
MYA 5/2Y/F1 |
 |
 |
 |
| Max capacity |
0,8/3 g |
5
g |
| Readability |
1/10 µg |
1 µg |
| Tare range |
-3 g |
-5 g |
| Repeatability * |
1 µg |
1,6 µg |
| Linearity |
±3 µg |
±5 µg |
| Eccentric load deviation |
3 µg |
5 µg |
| Sensitivity offset |
1,5×10-6×Rt |
| Sensitivity temperature drift |
1×10-6/°C×Rt |
| Sensitivity stability |
1×10-6/Year×Rt |
| Minimum weight (USP) |
3 mg |
4,8 mg |
| Minimum weight (U = 1%, k = 2) |
0,2 mg |
0,32 mg |
| Pan size |
ø 16 + ø 60 mm
(weighing pan for filters) |
ø100 mm |
ø160 mm |
| Weighing chamber dimensions |
ø 90×90 mm |
ø118×35 mm |
ø168×35 mm |
| Stabilization time |
5 s |
| Adjustment/calibration |
internal
(automatic) |
| Power supply |
110 ÷ 230 V AC / 50 ÷ 60 Hz / 13,5 ÷ 16 V DC / 1,1 A |
| Display |
5,7'' touch screen |
| Interface |
2×USB, 2×RS 232, Ethernet, 2Inputs/2Outputs |
| Net weight/Gross weight |
10,2/14,7
kg |
| Packaging size |
565×565×355
mm |
| Ambient conditions: |
| Working temperature |
+10° ÷ +40°C |
| Change rate of working temperatures |
±0,3 °C/h (±1 °C/8h) |
| Atmospheric humidity |
45% ÷ 65% |
| Change rate of atmospheric humidity |
±1%/h (±4%/8h) |
|